Description

 

The device image is for illustrative purposes only. The device will be manufactured according to the technical specifications.

The ultrasonic processor is one of the commonly used pieces of equipment for laboratory sample pre-processing.

It is widely used in nano-industry, biology, chemistry, pharmaceuticals, paints, optics, jewelry, aerospace, hardware, automobile manufacturing, and other fields.

Ultrasonic homogenizers are multi-functional and multi-purpose devices that use intense high-frequency sound to trigger cavitation in the liquid.

As cavitation bubbles collapse, high shear energies disrupt animal and plant tissue and break down yeast, bacteria, and spores.

The ultrasonicator can also be used for non-biological applications such as emulsification, nanoparticle dispersion, intensive washing, or accelerating chemical reactions (sonochemistry).

TECHNICAL SPECIFICATIONS

2.1. Digital LCD display

2.2. Fully microprocessor controlled and fully programmable

2.3. Automatic adjustment feature for ease of use and optimum processing efficiency

2.4. Temperature display and control device feature

2.5. 2.6. Automatic amplitude adjuster must be present. Ability to measure amplitude change upon delivery of the device.

2.7. Power spread indicator for accuracy and repeatability, Variable power output.

2.8. Integrated Noise Reduction Chamber to reduce cavitational sound emitted during sonication.

2.9. Ability to mix, emulsify and homogenize solutions.

2.10. Ability to reduce the particle size of suspensions.

2.11. Ability to disperse suspensions.

2.12. Ability to catalyze chemical reactions.

2.13. Frequency must be at least 22.6 kHz and have automatic frequency monitoring capability. Ability to measure frequency upon delivery of the device.

2.14. Output power: 165 W (must be continuously adjustable from 0-100%).

2.15. Length of the probe sonicator's piezoelectric holder: 100mm

2.16. Probe resonance frequency should be adjustable between 18 kHz and 35 kHz.

2.17. Sonicator device's industrial registration certificate

2.18. Total operating timer: 1 – 99s with pause feature

2.19. Ultrasonic output pulse: Ultrasonic on timer: 1 sec – 99 min

2.20. Ultrasonic off timer: 1 sec – 99 min

2.21. Power ratio should be adjustable between 0% and 100% on the LCD screen.

2.22. Ultrasonic probe diameter should be at least 6 mm and length at least 60 mm.

2.23. Sample processing volume: between 10 ml and 300 ml

2.24. The device must have a 2-year warranty.

2.25. The device must be delivered within a maximum of 2 weeks.

2.26. Within 1 week after the device delivery, at least 3 people must receive at least 5 hours of device maintenance and practical usage training.


 

SOCIAL MEDIA